The powder you weighed is not the powder you think you weighed

Bas Millenaar , LabTIE International B.V.


A balance is the most trusted instrument on the bench. It is calibrated, it is checked, and its reading goes into the record without a second thought. That trust is well placed as far as it goes: the balance reports the mass on its pan correctly.

The balance is right, the powder changed

What it cannot report is what that mass consists of. Weigh a hygroscopic solid in a room at 45 percent relative humidity and part of the number on the display is water that was not in the container when you opened it. The reading is accurate and the assumption behind it is wrong, which is the most durable kind of measurement error, because it survives every check aimed at the instrument.

The gradient nobody logs

The effect is easy to dismiss as a rounding matter until you look at a 96-well plate being filled by hand. Filling a plate takes minutes, the container stands open throughout, and the material in it is taking up water the whole time. The portion that goes into well A1 and the portion that goes into well H12 are therefore not the same material. Same label, same lot number, same nominal mass, and a different water content.

That difference does not present itself as an error. It presents itself as a gradient across the plate, running in the order in which the wells were filled. A dose-response curve laid out from low to high across the rows will carry it. A screening plate will show a drift that gets attributed to evaporation, to edge effects, or to the reader.

The point is not that the drift is large. It is that it is systematic, invisible in the run record, and indistinguishable afterwards from the effects it is usually blamed on.

What the guidance covers, and what it does not

The two documents an analyst reaches for cover the ends of the problem and not the middle. USP General Chapter <1251>, Weighing on an Analytical Balance [2], is written from the instrument outward: minimum weight, repeatability, measurement uncertainty, the environment the balance needs. It tells you what the instrument can do.

The pharmacopoeial classification is written from the material outward. The European Pharmacopoeia sorts substances by the mass gained after 24 hours at 25 degrees Celsius and 80 percent relative humidity [3]: slightly hygroscopic from 0.2 up to 2 percent, hygroscopic from 2 up to 15 percent, very hygroscopic above 15 percent, and deliquescent when the material takes up enough water to form a liquid. It tells you what the material is.

Neither answers the question the bench poses: what happens to this material in the four minutes my container stands open, in a room whose humidity I neither control nor record. There is no chapter for that interval, and in most laboratories no number either.

The physics is not in doubt. Every solid has an equilibrium relative humidity, and the reference values that define the humidity scale itself were established over saturated salt solutions by Greenspan in 1977 [1]: about 11 percent for lithium chloride at 25 degrees Celsius, about 75 percent for sodium chloride. A bench at 45 percent sits above the equilibrium point of a great many materials, so the gradient is permanent and the uptake continuous for as long as the container stands open.

Why the balance cannot warn you

Three properties make this error hard to catch. It is invisible: the powder does not change colour or clump, the display simply reads a little higher than a minute earlier. It is one-directional: uptake only adds mass, so the true content of a portion is always lower than stated, and errors that can go either way average out across a plate while this one does not. And it is correlated with time and therefore with position, so replication does not dilute it but reproduces it, because every plate is filled in the same order, in the same way, by the same person.

Measuring it on your own bench

The measurement costs an afternoon and needs nothing the laboratory does not already own.

Use a balance with 0.01 mg readability, levelled and with the draught shield closed. Put a calibrated hygrometer and thermometer next to it and log temperature and humidity at the start and end of every run; those two numbers are part of the result, not background information. Dry the material over desiccant to constant mass, defined as two weighings 30 minutes apart agreeing within 0.05 mg.

Weigh about 200 mg into an open vial of fixed geometry and record the inner diameter and the fill height, because the exposed surface area drives the rate. Leave the vial on the pan and record the mass every 30 seconds for 30 minutes without touching it, five times per material, spread over at least three days, so that the variation in room humidity sits inside the data rather than outside it. Run a non-hygroscopic control alongside, glass beads for instance: it carries the same buoyancy, drift and electrostatic effects and none of the water, so it sets the zero line. Subtract it.

Two numbers come out. The slope over the first five minutes is the operationally relevant one, because that is the window in which real weighing happens. The total at 30 minutes can be compared with the published classification.

An observation from our own bench

We ran exactly this test on calcium chloride, a deliquescent salt, in an open vial on an ordinary laboratory bench. The mass rose by about 2 percent in the first eleven minutes and by about 4.75 percent after forty-one, still climbing at the end of the run. Dosed instead in a step that keeps the material closed, the same salt stayed within a few tenths of a percent for the whole period (Figure 1).

Mass gain of calcium chloride in an open vial
Fig.1: Mass gain of calcium chloride in an open vial on a laboratory bench, against the same salt dosed in a closed step. The markers are measurements; the line between them is drawn. In-house observation, single run, room humidity not recorded.

Three things have to be said alongside that figure. Calcium chloride is a worst case and not a typical powder: its equilibrium relative humidity is in the region of 30 percent at 25 degrees Celsius, so a bench at 40 to 50 percent never lets it equilibrate and the uptake simply continues. A mildly hygroscopic material will show a fraction of this. Second, it is a single run rather than a validated comparison. And third, the room humidity during that run was not recorded, which is precisely the omission this article is arguing against; the measurement predates the habit it recommends.

What the figure does show is the shape of the problem: the curve is steepest at the start, which is the part of the interval in which weighing actually happens, and it does not flatten within the time it takes to fill a plate.

What to do once you have the number

setup at weighing station during a measurement
Fig.2: The setup at the weighing station during a series of
measurements: an open container on the pan,
with a hygrometer visible in the picture.
Shorten the time the container stands open, and measure how long that actually is rather than estimating it. Work by difference from a closed container. Fill the hygroscopic members of a series consecutively, so that exposure is constant across the plate rather than increasing through it. Record room humidity next to every weighing session. And where the same portion is repeated across a plate, any dosing step that closes in seconds rather than minutes removes most of the exposure by itself (Figure 2).

None of this eliminates the effect. It makes it small, constant and recorded, which is what a method needs.

The LabTIE dispenser doses by volume rather than by mass. A plate with cavities of a fixed volume is filled from the powder container, a scraper levels it off, and then you pull the slider (Figure 3).
The LabTIE dispenser
Fig.3: The LabTIE dispenser enables fast and
accurate filling of 96-well plates

This opens the holes and all 96 doses drop into the wells in one movement. A very accurate result, as long as the powder is homogeneous. The cavity volume sets the dose, in the way a pipette tip sets a liquid volume, so a different dose means a different plate.

Where this does not matter

It does not matter for materials that are not hygroscopic, and most are not, nor when the portion is large and the tolerance loose, nor in a weighing room whose humidity is genuinely controlled rather than merely labelled. It disappears entirely if the work can be done from a solution instead of a solid.

It matters in the case that keeps recurring: a 96-well plate filled by hand on an ordinary bench, by someone who was told the balance reads to 0.01 mg and reasonably concluded that weighing was the step they did not need to worry about.

References

  1. Greenspan, L. (1977). Humidity fixed points of binary saturated aqueous solutions. Journal of Research of the National Bureau of Standards, 81A(1), 89 bis 96.
  2. United States Pharmacopeia, General Chapter <1251>, Weighing on an Analytical Balance.
  3. European Pharmacopoeia (Ph. Eur.)